SPECIFICATION

Parameters

21 reportable Parameters: WBC, Lym#, MID#, GRAN#, LYM%, MID%,GRAN%, RBC, HGB, HCT, MCV,MCHC, RDW-CV,RDW-SD, PLT,MPV,PDW, PCT,P-LCR, P-LCC

Performance
Parameter Precision Linear Range Contamination Rate
WBC < 2.0% (3.5-15.0x10⁹)/L (0-100.0x10⁹)/L _0.5% < 0.5%
RBC < 1.5% (3.0-6.0x10¹²)/L (0.10-8.00x10¹²)/L < 1.0% (70-120)fL < 0.5%
HGB < 1.5% (100-180)g/L (0-250)g/L - < 0.6%
MCV < 1.0% (70-120)fL< - (0.1000x10⁹)/L -
PLT < 4.0% (150-500x10⁹)/L (0.1000x10⁹)/L - < 1.0%
Sample

Predilution Mode   20 L

Whole Blood Mode   9 L

Efficient

60 Samples per hour

Storage

Up to 50,000 results including histograms, and patiet information

Compatibility

Build-in thermal printer External laser or inkjet printer 4 USB, 1 LAN support protocol Hl7 and compatible with LIS system

Environment

Working Voltage:   Ac100V 240V, 50 60Hz
Rated Power:   40-60VA
Working Temperature:   10°C 40°C
Relative Humidity:   20%-85%

Power Requirement

100v-200v
< 300VA
50Hz/60Hz

Reagent

Diluent 20L Build-in LYSE
Probe Cleanser

Display

10.4 inch colorful touchscreen

Dimension and weight

Dimension: Depth(400 mm)x width(285 mm)
x height (400 mm)
Weight: _20 Kg

hematology
Hemat-4 Plus
3-Part Hematology Analyzer
Download Hematology Analyzer Brochure

Emergency hotline

(+91) – 9711838655

We provide 24/7 customer support. Please feel free to contact us
for emergency case.